Abstract

A position-sensitive neutron detector has been developed for use in nuclear physics research. The detector consists of a ∅5.5 cm×100 cm long quartz tube filled with liquid scintillator viewed from both ends by photomultipliers and enclosed in a light-tight titanium container. The properties of the detector were determined both experimentally and by Monte Carlo simulations (EFEN code). A time resolution of 0.4 ns was reached resulting in the position resolution of less than 4 cm. The neutron registration efficiency varies from 36% to 20% within neutron energy range 1–10 MeV and is practically independent of the position along the detector length. Good n–γ separation is achieved for neutron energies greater than 0.5 MeV.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.