Abstract

Abstract Experimental investigations were carried out in order to identify the portable device requirements for in-vitro muscle tissue monitoring over time. Based on these investigations, specifications for the measurement system were defined considering the type and amplitude of excitation signals (500 mV), frequency range (1 kHz–10 MHz) and impedance range (10 Ω–4 kΩ). For these requirements, a portable device structure is proposed based on the magnitude ratio and phase difference detection using a gain phase detector. To fulfill the requirements of the gain phase detector circuit's inputs, an interface circuit is proposed with an error lower than 0.09% from 2 mV up to 200 mV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.