AbstractPolythymine oligonucleotide (PTO)‐modified gold electrode (PTO/Au) was developed for selective and sensitive Hg2+ detection in aqueous solutions. This modified electrode was prepared by self‐assembly of thiolated polythymine oligonucleotide (5′‐SH‐T15‐3′) on the gold electrode via AuS bonds, and then the surface was passivated with 1‐mercaptohexanol solution. The proposed electrode utilizes the specific binding interactions between Hg2+ and thymine to selectively capture Hg2+, thereby reducing the interference from coexistent ions. After exchanging the medium, electrochemical reduction at −0.2 V for 60 s, voltammetric determination was performed by differential pulse voltammetry using 10 mM HEPES; pH 7.2, 1 M NaClO4 as supporting electrolyte. This electrode showed increasing voltammetric response in the range of 0.21 nM Hg2+, with a relative standard deviation of 5.32% and a practical detection limit of 60 pM. Compared with the conventional stripping approach, the modified electrode exhibits good sensitivity and selectivity, and is expected to be a new type of green electrode.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.