Abstract

A voltammetric sensor has been developed for the individual as well as simultaneous determination of xanthine (XA) and hypoxanthine (HX) based on an electroactive-polymerised layer of para toluene sulphonic acid and gold nanoparticles composite modified glassy carbon electrode ([p(PTSA)]/AuNPs/GCE)]. Under optimized conditions, an enhancement in the oxidation currents with well-separated and well-resolved peak position and a lower shift in the peak potentials were observed. By square wave voltammetry, the simultaneous determinations of XA and HX were achieved in the linear ranges 6.00 × 10-4M to 3.00 × 10-6M and 5.00 × 10-4M to 1.00 × 10-5M with detection limits of 4.09 × 10-7M and 4.10 × 10-7M, respectively. The mechanistic aspects were unveiled from linear sweep voltammetric studies and found that the electrode processes were diffusion-controlled. Finally, the sensor was successfully employed for the simultaneous determination of spiked amount of XA and HX in synthetic urine and serum samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.