Abstract
Poly(vinyl chloride) (PVC) membrane electrodes based on bis(dialkyldithiophosphates) carriers were prepared and tested for selective detection of silver ions. A suitable lipophilicity of the carrier and appropriate co-ordination ability were found to be essential for designing an electrode with good response characteristics. A PVC membrane with 5% bis(diethyldithiophosphates) ionophore and 65% o-NPOE plasticizer satisfies these requirements. This electrode exhibits linear response over a concentration range of 10 −1–10 −6 mol l −1 Ag + with a slope of 57.3 mV/dec. This is a fast and simple analytical method, and it has been applied successfully in the determination of silver ions in electroplating wastewater.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.