Abstract

Metoclopramide (MCP) ion selective electrodes based on metoclopramide-phosphotungstic acid (MCP-PT) ion pair complex in PVC matrix membrane were constructed. The plasticizers used were tri-butyl phosphate (TBP), di-octyl phenyl phosphonate (DOPP), di-butyl phthalate (DBPH), di-octyl phthalate (DOP), di-butyl phosphate (DBP), bis 2-ethyl hexyl phosphate (BEHP). The sensors based on TBP, DOPP, DBPH and DOP display a fast, stable and linear response with slopes 59.9, 57.7, 57.4, 55.3 mV/decade respectively at pH ranged 2-6. The linear concentration range between 1.0×10-5 – 1.0×10-2 M with detection limit 3.0×10-6 and 4.0×10-6 M for electrodes using TBP, DOPP and DBPH while electrode using DOP shows a linear concentration range between 3.0×10-5 – 1.0×10-2 M and detection limit 6×10-6 M. The selectivity coefficient values were calculated for MCP with respect to different inorganic cations, sugars and amino acids using the MCP depending on TBP plasticizer which is also used for the determination of MCP in pure solutions and in pharmaceutical preparations by direct, standard addition methods and as an indicator electrode in potentiometric titration with phosphotungstic acid (PT) giving satisfactory results.
 Key words: Metoclopramide-HCl membrane sensors , Metoclopramide determination , pharmaceutical analysis , PVC membrane sensors

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.