Abstract

In this paper, we report on the optimization of the shape of dry microneedle electrodes for electroencephalography (EEG) on hairy locations and compare the electrodes we developed with conventional wet electrodes. We propose the use of SU-8-based candle-shaped microneedle electrodes (CMEs), which have pillars of 1.0 mm height and 0.4 mm diameter with a gap of 0.43 mm between pillars. Microneedles are formed on the top of the pillars. The shape was determined by how well the pillars can avoid hairs and support the microneedles to penetrate through the stratum corneum. The skin–electrode contact impedances of the fabricated CMEs were found to be higher and less stable than those of conventional wet electrodes. However, the CMEs successfully acquired signals with qualities as good as those of conventional wet electrodes. Given the usability of the CMEs, which do not require skin preparation or gel, they are promising alternatives to conventional wet electrodes.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call