Abstract

AbstractAppropriately formulated polymeric membrane electrodes are shown to exhibit analytically useful nonequilibrium potentiometric responses to biologically important polyanions with high phosphate content, including polyphosphates and nucleic acids, in the μg/mL concentration range This response is dependent on the charge density, lipophilicity, and molecular weight of the polyanion species. The application of these electrodes for monitoring enzymatic activities is also demonstrated, using acid phosphatase/polyphosphate as a model system. The detection limit of this assay was found to be 0.9 ± 0.4 mU/mL with a linear range of 5–100 mU/mL. A similar electrochemical assay scheme can be utilized to monitor ribonuclease activity, using a RNA homopolymer as the substrate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.