Abstract
Glenoid component stability is essential to ensure successful long-term survivability following total shoulder arthroplasty. As such, this computational study assessed the stability of five all-polyethylene glenoid components (Keel, Central-Finned 4-Peg, Peripheral 4-Peg, Cross-Keel, and Inverted-Y), using simulated joint loading in an osteoarthritic patient cohort. Stability was assessed on the basis of component micromotion in the tangential and normal directions. Maximum tangential micromotion occurred in the Cross-Keel (146 ± 46 µm), which was significantly greater (p < .001) than the other components. Maximum normal micromotion occurred in the Inverted-Y (109 ± 43 µm), which was significantly greater (p ≤ .002) than the other four components. In general, the Central-Finned 4-Peg exhibited the least normal and tangential micromotion, while the keeled components shown the highest normal and tangential micromotion. This study suggests that modifications to keeled designs do not improve component stability under the conditions tested, and pegged components show superior computational stability.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Computer Methods in Biomechanics and Biomedical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.