Abstract
In a series of 111 AGC total knee arthroplasties, eight failures of the polyethylene part of the metal-backed patellar component were seen after a period of 7-22 months. All of these patients experienced a characteristic grating sound as the knee was bent, and skyline radiographs revealed direct metal-on-metal contact. At revision, fracture and separation of the polyethylene part were found together with an extensive, black, thickened synovium and abrasive wear of the naked metal backing of the patellar component. The cemented metal backings were all solidly anchored in the patella. Based on this high failure rate, we have abandoned the metal-backed patellar component in favor of an all-polyethylene component.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.