Abstract

Both polyclonal and monoclonal antibodies have been raised against purified human fibroblast tissue inhibitor of metalloproteinases (TIMP) and characterised. Combinations of antibodies were tested for their suitability in a double antibody sandwich ELISA to measure TIMP. Two combinations were applicable to the immunoassay: (i) a monoclonal capture antibody with a polyclonal detecting antibody; (ii) two monoclonal antibodies. The assay range was (i) 2-50 ng/ml and (ii) 5-50 ng/ml of human TIMP. The levels of TIMP in several human body fluids were measured using assay (ii), and values in ng/ml with standard deviations (n-1) obtained as follows: serum, 299 ± 62; plasma, 109 ± 35; amniotic fluid, 724 ± 62; cerebrospinal fluid, 144 ± 156; saliva, 209 ± 103.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call