Abstract

Polyaniline nanotubes (PANI-NT) based film electrophoretically deposited onto indium–tin–oxide (ITO) coated glass plate has been utilized for covalent immobilization of lipase (LIP), via glutaraldehyde (Glu), for triglyceride detection using impedimetric technique. It is shown that fatty acid molecules produced during triglyceride hydrolysis result in change in charge transfer resistance ( R CT) of PANI-NT film with varying triglyceride concentration. LIP/Glu/PANI-NT/ITO bioelectrode has linearity as 25–300 mg dL −1, sensitivity as 2.59 × 10 −3 KΩ −1 mg −1 dL, response time as 20 s and regression coefficient as 0.99. A low value of apparent Michaelis–Menten constant (∼0.62 mM) indicates high enzyme affinity to tributyrin. The LIP/Glu/PANI-NT/ITO bioelectrode has been utilized to estimate triglyceride in serum samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.