Abstract

Electrocardiography (ECG) has been instrumental for early detection of cardiovascular anomalies. In this research, we successfully prototyped and evaluated the performance of a novel dry electrode as ECG sensor. Copper was selected as dry electrode material due to its good balance between conductivity and affordability. Polyaniline was used as a conductive coating to facilitate the conversion of ionic currents from the human body into electrical currents in electronic circuits. The coating was carried out via electrodeposition technique. Optimal electrodeposition time of 20 min using acetic acid as a dopant was established. This yields in the dry electrode with comparable performance to the certified wet electrode currently available in the market, shown by PQRST signal correlation between dry electrodes (this study) and the wet electrode (benchmark) which is close to zero. The dry electrode prototyped in this study is characterized to have a conductivity of 7 x 10 -4 S/mm, a resistance of 10 Ω, capacitive reactance of 140 Ω, and excellent signal stability showing a value of 256.5 μV consistently for 2 hours.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.