Abstract

The performance of a thin-layer flow detector with a glassy carbon electrode coated with a film of protonated poly(4-vinylpyridine) is described. Substantial improvement in the selectivity of amperometric detection for liquid chromatography and flow injection systems is observed as a result of excluding cationic species from the surface. The detector response was evaluated with respect to flow rate, solute concentration, coating scheme, film-to-film reproducibility, and other variables. Despite the increase in diffusional resistance, low detection limits of ca. 0.04 and 0.10 ng of ascorbic acid and uric acid, respectively, are maintained. Protection from organic surfactants can be coupled to the charge exclusion effect by using a bilayer coating, with a cellulose acetate film atop the poly(4-vinylpyridine) layer. Applicability to urine sample is demonstrated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.