Abstract

This work presents polarization-sensitive optical coherence tomography (PS-OCT) using only linearly polarized light. The main characteristics of linear PS-OCT include the following. (1) In a turbid medium with linear birefringence, the degree of polarization (DOP) of backscattered linearly incident polarized light does not change significantly with the increase of birefringence value; (2) the phase retardation of the sample can be obtained from only the phase difference between the orthogonal polarization heterodyne signals, so a common phase noise rejection mode is introduced. These advantages may not only support the immunity of background phase noise and laser frequency noise, but also increase the sensitivity of phase retardation measurements; (3) the dynamic range of phase retardation in this system is between 0° and 180°

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.