Abstract
Single-molecule localization microscopy has been widely applied to count the number of biological molecules within a certain structure. The percentage of molecules that are detected significantly affects the interpretation of data. Among many factors that affect this percentage, the polarization state of the excitation light is often neglected or at least unstated in publications. We demonstrate by simulation and experiment that the number of molecules detected can be different from -40 up to 100% when using circularly or linearly polarized excitation light. This is determined mainly by the number of photons emitted by single fluorescent molecule, namely the choice of fluorescence proteins, and the background noise in the system, namely the illumination scheme. This difference can be further exaggerated or mitigated by various fixation methods, magnification, and camera settings We conclude that the final choice between circularly or linearly polarized excitation light should be made experimentally, based on the signal to noise ratio of the system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.