Abstract

The incorporation of polarization sensitivity into optical coherence tomography (PS-OCT) imaging can greatly enhance utility by allowing differentiation via intrinsic contrast of many types of tissue. In fiber-based OCT systems such as those employing endoscopic imaging probes, however, polarization mode dispersion (PMD) can significantly impact the ability to obtain accurate polarization data unless valuable axial resolution is sacrificed. In this work we present a new technique for compensating for PMD in endoscopic PS-OCT with minimal impact on axial resolution and without requiring mutually coherent polarization inputs, needing only a birefringent structure with known orientation in view (such as the catheter sheath). We then demonstrate the advantages of this technique by comparing it against the current state of the art approach.