Abstract

Abstract Reliable results when fitting X-ray data require proper consideration of the statistics involved. We probe the impact of Gaussian versus Poisson statistics at low count levels using both the standard χ2 method and maximum likelihood based on Poisson (C) statistics. The difference is studied and quantified through simulated spectra with known properties. We then test the results through analysis of Mn Kα calibration data taken with the flight spare microcalorimeter for the Hitomi soft X-ray spectrometer. Through comparison with simulations, our results show that the χ2 method tends to give overly optimistic estimates of the detector energy resolution, in particular when there are few counts. Given an energy resolution of ∼5 eV and a line with about 100 photons, the line width becomes ∼10% lower in the χ2 method than in Poisson statistics. This is a consequence of the uncertainties being dominated by counting statistics, and therefore highlights the need to choose the appropriate fit statistic.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.