Abstract

As the key component of intravascular ultrasound (IVUS) imaging systems, traditional commercial side-looking IVUS transducers are flat and unfocused, which limits their lateral resolution. We propose a PMN-PT single crystal IVUS transducer with a half-concave geometry. This unique configuration makes it possible to conduct geometric focusing at a desired depth. To compare performances, the proposed and the traditional flat transducer with similar dimensions were fabricated. We determined that the half-concave transducer has a slightly higher center frequency (35MHz), significantly broader -6dB bandwidth (54%) but a higher insertion loss (-22.4dB) compared to the flat transducer (32MHz, 28%, and -19.3dB, respectively). A significant enhancement of the lateral resolution was also confirmed. The experimental results are in agreement with the finite element simulation results. This preliminary investigation suggests that the half-concave geometry design is a promising approach in the development of focused IVUS transducers with broad bandwidth and high lateral resolution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.