Abstract

Plume generated by electrosurgical techniques is a health hazard to patients and dermatologists. To compare the particle concentration generated by various energy devices used in dermatologic surgery. Five surgical techniques were tested on human tissue samples in a closed chamber. A particle counter, positioned at a fixed point 20 cm away from the sample, recorded the concentrations of aerosolized particles generated over 7 particle sizes (0.3, 0.5, 0.7, 1, 2.5, 5, and 10 μm). Monopolar electrocoagulation created the greatest concentration of particles followed by electrocautery, electrodesiccation, electrofulguration, and bipolar electrocoagulation. Bipolar electrocoagulation created 80 times fewer 0.3 μm particles and 98 times fewer 0.5 μm particles than monopolar electrocoagulation. Across all electrosurgical techniques, the greatest concentrations of particles generated were of the 0.3 and 0.5 μm particle size. Bipolar electrocoagulation created the lowest concentration of particulate matter. Given the noxious and hazardous nature of surgical plume, the bipolar forceps offer surgeons a safer method of performing electrical surgery for both the surgical staff and the patient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call