Abstract

Thirty-one units of hospital blood bank apheresis PLTs were obtained. PLT-rich plasma (PRP) aggregometry and thromboelastography (TEG) were performed on the unit samples before and after the units were infused through a Ranger blood/fluid warming device. There were no differences in any of the aggregometry results before and after infusion of the PLTs through the blood warmer (all P > .32). There was a significant reduction in the TEG maximum amplitude (MA) of 69.8 ± 7.9 mm before and 66.0 ± 8.8 mm after (P < .001) infusion of the PLTs through the blood warmer and α angle 61.8 ± 9.4° before and 59.3 ± 8.2° after (P = .044) infusion of the PLTs through the blood warmer, although both mean values were within normal range for the TEG and not clinically significant. There were very good correlations of aggregometry and TEG results before and after infusion of the PLTs through the blood warmer device. This study did not demonstrate significant deleterious effect on PLT function from infusing apheresis PLT units through a blood warming device by PRP aggregometry. We did detect a statistically significant-but not clinically significant-reduction in TEG MA and α angle. The prohibition of transfusing PLT units though the Ranger blood warming device is not indicated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.