Abstract
AFO is a device, which is usually recommended for hemiplegics to help their pathologic gait. To date, no good engineering guidance is available. We analyze mechanical characteristics of AFO and evaluate structural stability effect of thickness of ankle foot orthosis (AFO). The study is to provide a guidance to make a functional AFO. Basic material characteristics such as Young’s modulus and Possion’s ratio were measured through material tests. Finite element models were made and analyzed based on various shapes and gait analysis. Young’s modulus of polypropylene is 1.15 GPa and Possion’s ratio is 0.35. Rigidity of AFO is more affected by the ankle width than the thickness of ankle area. The strain calculated by finite element analysis (FEA) in the neck area is twice as much as experimental results but the both of their trend show similar pattern. AFO was studied through its structural and functional analysis. Experimental structural analysis and FEA were made and compared.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.