Abstract

The development of highly sensitive electrochemiluminescence (ECL) immunosensors by using functional nanoparticles as signal amplifiers is a solution towards sensitive determination of many low concentration disease biomarkers. Herein, a sensitive aptamer-based, sandwich-type surface plasmon enhanced electrochemiluminescence (SPEECL) immunosensor was demonstrated for the detection of cardiac troponin I (cTnI), by means of aptamer conjugated CdS QDs and AuNPs as ECL luminophores and plasmon sources, respectively, in which Tro4 aptamer was used as a capture probe for cTnI and Tro6 aptamer as a detecting probe. The signal of the developed SPEECL system showed ~ 5-fold increment as compared to that of without AuNPs. Using this ECL platform for the detection of cTnI, a linear range and the limit of detection (LOD) were found to be 1 fg/mL - 10 ng/mL and 0.75 fg/mL, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.