Abstract

Double-pulse Laser-Induced Breakdown Spectroscopy (LIBS) in an orthogonal configuration was used to investigate plasma temperature and electron density effects on Mg II emission spectral line broadening. The experiments were carried out with two Nd:YAG lasers, one operating at 355 nm for ablation and the other one at 1064 nm for plasma reheating in air at atmospheric pressure. Temporally resolved plasma temperature and electron density were measured at various delay times. Data in this study show prolonged emission of Mg II (280.27 nm) as well as enhancement of the signal intensity when using double-pulse excitation compared to the single-pulse case. An enhancement of ∼8× was attained with a delay between the laser pulses equal to 1 μs. The enhancement was accompanied by higher plasma temperature and increased electron density. The double-pulse LIBS configuration provides energy to sustain the plasma emission at a period in time when the linewidth is minimum, thereby improving the analytical capabilities of low spectral resolution instrumentation typically used in LIBS system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call