Abstract

Liquid chromatography-mass spectrometry (LC-MS) is widely used in the detection of pesticide residues. However, the detection sensitivity of low-polarity pesticides by commonly used electrospray ionization may be severely suppressed, which greatly affects the limit of detection and repeatability. Herein, a plasma-excited nebulizer gas-assisted electrospray ionization (PENG-ESI) device has been developed. By introducing the discharge plasma formed by Tesla coil into the electrospray nebulizer gas channel, the sensitivity of low-polarity pesticides was significantly increased while maintaining sensitivity to polar pesticides. Under the optimized conditions, the limit of detection for S-bioallethrin was achieved at the level of 100 pg/g with good linearity (R2 > 0.99) and precision (RSD ≤ 4.61%). The matrix effect of a series of spiked matrix samples is less than 13.1%. Finally, different pyrethroid pesticide residues were successfully analyzed without separation, highlighting that the technology has potential application prospects in food quality control, environmental monitoring, and other fields.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.