Abstract

The use of the colorimetric Jaffé method for the measurement of creatinine in mouse and rat plasma has been criticized as prior studies have shown a dramatic overestimation. We compared a colorimetric picric acid, an enzymatic, and a high-performance liquid chromatography (HPLC) method to assess their appropriateness for routine measurements of creatinine in plasma of healthy and diseased mice (n=61) and rats (n=56). For the colorimetric Jaffé method a pronounced overestimation is confirmed. Additionally the method showed interference with hemoglobin already in a very low, non-visible concentration range in rat plasma. The enzymatic measurement demonstrated a hemoglobin interference in mice, only when hemolysis was visible. The comparison between HPLC and the enzymatic measurement gave a good agreement between both methods in both species. Therefore the enzymatic method fulfills the requirements for a routine screening test for plasma creatinine in healthy as well as diseased mice and rats Kiover a broad concentration range.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.