Abstract

Better sensitivity and interface of ambient sampling/ionization mass spectrometry remain a challenge. Herein, a novel, plasma-based, ambient sampling/ionization/transmission (PASIT) integrated source in a pin-to-funnel configuration was developed for the sensitive analysis of complex samples. With the funnel sleeve directly affected by direct-current discharge plasma, PASIT combines the ability to sample/ionize analyte molecules and then efficiently collect/transport charged mass species under atmospheric pressure and consequently shows an improved sensitivity. The integrated source enhances the signal intensity by more than 2 orders of magnitude compared with the previous pin-to-plate plasma source without significant background addition. A surface limit of detection (LOD) of 130 fmol mm-(2) (S/N = 3) has been achieved for clenbuterol on filter paper with an argon carrier gas. Demonstrated applications include the direct determination of active ingredients from drugs and symbolic compounds from natural plants and cholesterol from mouse brain tissue sections.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.