Abstract

We have conducted laboratory experiments as an analogue to planetary XRF (X-ray fluorescence) missions in order to investigate the role of changing incidence (and phase) angle geometry and sample grain-size on the intensity of XRF from regolith-like samples. Our data provide evidence of a grain-size effect, where XRF line intensity decreases with increasing sample grain-size, as well as an almost ubiquitous increase in XRF line intensity above incidence angles of ∼60°. Data from a lunar regolith simulant are also used to test the accuracy of an XRF abundance algorithm developed at the Rutherford Appleton Laboratory (RAL), which is used to estimate the major element abundance of the lunar surface from Chandrayaan-1 X-ray Spectrometer (C1XS) XRF data. In ideal situations (i.e., when the input spectrum is well defined and the XRF spectrum has a sufficient signal to noise ratio) the algorithm can recover a known rock composition to within 1.0elemental wt% (1σ).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.