Abstract

A comparative study at 2.45 GHz concerning both measurement and reconstruction parameters for planar and cylindrical configurations is presented. For the sake of comparison, a numerical model consisting of two nonconcentric cylinders is considered and reconstructed using both geometries from simulated experimental data. The scattered fields and reconstructed images permit extraction of very useful information about dynamic range, sensitivity, resolution, and quantitative image accuracy for the choice of the configuration in a particular application. Both geometries can measure forward and backward scattered fields. The backscattering measurement improves the image resolution and reconstruction in lossy mediums, but, on the other hand, has several dynamic range difficulties. This tradeoff between forward only and forward-backward field measurement is analyzed. As differential temperature imaging is a weakly scattering problem, Born approximation algorithms can be used. The simplicity of Born reconstruction algorithms and the use of FFT make them very attractive for real-time biomedical imaging systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.