Abstract

We investigated the plan complexity of volumetric modulated arc therapy (VMAT) with knowledge-based plan (KBP) for oropharyngeal cancer (OPC) with a single optimization and whether it could be used clinically. KBP model was configured using 55 consecutive OPC and nasopharyngeal cancer plans. Plan complexity as a characteristic of multileaf collimator (MLC) motion and γ pass rate (2%/2 mm criterion) were compared between clinical manual plan (CMP) and KBP for other 10 plans. Plan complexity metrics that had significant differences (p<0.05) (CMP vs. KBP), were mean lateral displacement of MLC from central axis (15.82 mm vs. 18.90 mm), proportions of MLC aperture sizes of ≤5 mm (0.14 vs. 0.11), ≤10 mm (0.24 vs. 0.19), and ≤20 mm (0.41 vs. 0.34), and monitor units (578.68 vs. 505.04). The γ pass rate was 91.3% vs. 93.3%. Single optimized KBP for OPC had simple plan complexity features and comparable delivery accuracy to CMP, and could be clinically applied.