Abstract

Pulse oximetry is widely used to assess oxygen saturation (SpO2) in order to guide patient care and monitor the response to treatment. However, inappropriate oximeter probe placement has been shown to affect the measured oximetry values in healthy and normoxic outpatients. This study evaluated how treatment decisions might be impacted by SpO2 values obtained using a finger probe placed on the pinna of the ear in a cohort of 46 patients receiving non-invasive ventilation compared with values obtained from a probe on the finger and the results of arterial blood gas (ABG) (SaO2) analysis. Bland-Altman analysis was performed to evaluate agreement between the methods. Finger probe saturation was not statistically different from SaO2, with a mean difference of -0.66% (P>0.05). Saturation from the ear was significantly different (-4.29%; P<0.001). Subgroup analysis in hypoxic patients (SaO2<90%) showed a significant difference between ABG SaO2, and finger and ear SpO2. The study provides evidence that placement of a finger probe on the ear is unsafe clinical practice, potentially leading to patient mismanagement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.