Abstract

The RVG-4 permits automatic exposure compensation (AEC). The purpose of this investigation was to determine the effects of AEC on image contrast. Images were made either with or without a dental QA jaw phantom using a fixed image projection geometry. Exposures were 6.3 through 27.3 μC/kg using an X-ray generator operated at 70 kVp. Region of interest pixel value distributions were measured at tissue thicknesses in this phantom, and the average pixel values and signal-to-noise ratios (SNR) were calculated. The use of AEC without an object in place resulted in a disproportionate relationship between pixel value and exposure with a marked reduction in SNR. The use of AEC on under- and over-exposed images of the phantom simultaneously enhanced image contrast and reduced SNR. Thus, AEC provides a convenient and quick method for achieving high-contrast images with sub-optimal exposures, however, this could lead to inappropriate patient dosages if the function is used for over-exposed images. AEC reduces the SNR and produces disproportionate pixel values relative to exposure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.