Abstract

The chemical characterization of size-segregated samples is helpful to study the aerosol effects on both human health and environment. The sampling with multi-stage cascade impactors (e.g., Small Deposit area Impactor, SDI) produces inhomogeneous samples, with a multi-spot geometry and a non-negligible particle stratification.At LABEC (Laboratory of nuclear techniques for the Environment and the Cultural Heritage), an external beam line is fully dedicated to PIXE–PIGE analysis of aerosol samples. PIGE is routinely used as a sidekick of PIXE to correct the underestimation of PIXE in quantifying the concentration of the lightest detectable elements, like Na or Al, due to X-ray absorption inside the individual aerosol particles. In this work PIGE has been used to study proper attenuation correction factors for SDI samples: relevant attenuation effects have been observed also for stages collecting smaller particles, and consequent implications on the retrieved aerosol modal structure have been evidenced.