Abstract

We describe problems encountered in our first 136 patients, with 95 aneurysms, who underwent spiral CT for investigation of possible aneurysms involving the circle of Willis and adjacent major vessels, and who had surgical and/or angiographic confirmation. There were seven false-positive cases, of which the first three could be explained by operator inexperience. There were four false negatives, all small aneurysms; two were not seen because of operator error and two were hidden by an adjacent larger aneurysm. Clip artefacts prevented diagnostic studies in six of 21 postoperative studies. One aneurysm was outside the CT field of view, being on a pericallosal artery. One basilar artery tip aneurysm was excluded from the field of the CT study because of a planning error. Inspection of the axial source images is critical if the diagnosis of small or thrombosed aneurysms is to be made. Close attention to image acquisition and computer modelling is required to reduce errors in spiral CT angiography of intracranial aneurysms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.