Abstract

The accuracy of the arterial pressure-based cardiac output FloTrac-Vigileo system remains unacceptably low during haemodynamic instability. Data show that the measurement of cardiac output (CO) is strongly influenced by changes in factors that affect arterial blood pressure (ABP) - for example, vascular tone and compliance and the arterial site - independently of true changes in CO. Although in theory the autocalibration algorithm of FloTrac-Vigileo should adjust for those changes, the model undercompensates (or overcompensates) for prominent increases (or decreases) in vascular tone and compliance, making the system largely dependent on changes in ABP. These limitations make FloTrac-Vigileo accurate in stable haemodynamic conditions only, and until more robust algorithms and further validation studies become available, we should be aware that during haemodynamic instability or in extreme conditions of vasodilation or vasoconstriction, the measured CO may diverge from an independent bolus indicator dilution measurement, particularly if a peripheral artery is used. In these conditions, we advocate the use of transpulmonary indicator dilution via a femoral artery.

Highlights

  • The accuracy of the arterial pressure-based cardiac output FloTrac-Vigileo system remains unacceptably low during haemodynamic instability

  • There has been a trend toward the use, in intensive care units (ICUs) and in operating theatres, of ‘minimally invasive’ haemodynamic monitoring systems for the continuous measurement of cardiac output (CO)

  • Lifesciences LLC, Irvine, CA, USA) does perhaps deserve this epithet as it is designed to run from any arterial line and requires no calibration. This latter capability is a consequence of a sophisticated algorithm that the device employs to analyse the arterial pressure waveform (APW), whether obtained from the radial or the femoral artery, to determine the presumed non-linear proportionality between arterial blood pressure (ABP) and stroke volume (SV) and give an estimate of CO

Read more

Summary

Introduction

The accuracy of the arterial pressure-based cardiac output FloTrac-Vigileo system remains unacceptably low during haemodynamic instability.

Results
Conclusion