Abstract
Herein, a pipette-tip-enabled digital nucleic acid analyzer for high-performance COVID-19 testing is demonstrated. This is achieved by digital loop-mediated isothermal amplification (digital LAMP or dLAMP) using common laboratory equipment and materials. It is shown that simply fixing a glass capillary inside conventional pipette tips enables the generation of monodisperse, water-in-oil microdroplets with benchtop centrifugation. It is shown that using LAMP, the ORF1a/b gene, a standard test region for COVID-19 screening, can be amplified without a thermal cycler. The amplification allows counting of fluorescent microdroplets so that Poisson analysis can be performed to allow quantification with a limit of detection that is 1 order of magnitude better than those of nondigital techniques and comparable to those of commercial dLAMP platforms. It is envisioned that this work will inspire studies on ultrasensitive digital nucleic acid analyzers demanding both sensitivity and accessibility, which is pivotal to their large-scale applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.