Abstract
Despite a number of preventive mechanisms, inadvertent administration of nitrous oxide in place of oxygen can lead to fatal hypoxemia. Here we report two cases of hypoxia that occurred when we switched to the emergency cylinder for oxygen supply after exhaustion of the main oxygen cylinder. The urgency shown by the anesthetist and operating room staff to restore the main oxygen supply prevented any fatalities from occurring in our case. We found that there was incorrect painting of the nitrous oxide cylinder with the color code of oxygen. Further, damaged pins on the yoke assembly allowed the attachment of the faulty E cylinder to the machine. Even though such errors are made by the supplier we suggest that all equipment including the cylinder be thoroughly checked by the anesthetist. This also highlights the role of respiratory gas monitoring in the prevention of such mishaps.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.