Abstract

Objectives To determine if Electrode Voltage (EV) measurements are potentially suitable as a test for detecting extra-cochlear electrodes in cochlear implants (CIs). Methods EV measurements were made using surface electrodes in live mode in 17 adult cochlear implant (CI) users. Repeatability, the effects of stimulation level, CI active electrode position, (active) recording electrode position and stimulation mode (for Nucleus devices) were investigated. Results/discussion Recordings made in monopolar mode showed good repeatability when the active recording electrode was placed on the ipsilateral earlobe; voltages increased linearly with stimulation level as expected. EVs for basal electrodes differed greatly between partially inserted/migrated devices, fully inserted devices with all electrodes activated, and those with deactivated basal electrodes [χ 2(2) = 10.2, p < 0.05 for the most basal electrode]. EVs for Nucleus devices were small for electrodes on the array when compared to those for monopolar return electrodes, except for the participant with extra-cochlear electrodes. We argue that fibrosis around the electrode array facilitated current flow across the round window in this case. Conclusion The test appears to be a viable approach to detect electrode migration and extra-cochlear electrodes in adult CI users and may also be sensitive to discomfort caused by current leakage from the basal end of the cochlea.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.