Abstract

The use of a piezoelectric tube for the photoacoustic gas-phase determination of NO2 as a model analyte is demonstrated. The tube is made from lead zirconate titanate with 30 mm length and 5.35 mm internal diameter. Its inner and outer surfaces are coated with electrodes. The tube serves as both, resonance body and transducer. The design is thus simpler than the usual combination of resonance tube and microphone as the two functions are embodied in the same component. The main resonance frequency of the tube was found to be 5341 Hz. A blue laser diode emitting at 450 nm was employed as light source for the determination of NO2. The limit of detection was determined as 83 ppbV and the calibration curve was linear with a coefficient of determination (r2) of 0.9998 up to the highest concentration of 15 ppmV tested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.