Abstract

The piezoelectric shear gauge (PSG) [Christensen and Olsen, Rev. Sci. Instrum. 66, 5019 (1995)] is a rheometric technique developed to measure the complex shear modulus of viscous liquids near their glass transition temperature. We report recent advances to the PSG technique: (1) The data extraction procedure is optimized, which extends the upper limit of the frequency range of the method to between 50 and 70 kHz. (2) The measuring cell is simplified to use only one piezoelectric ceramic disk instead of three. We present an implementation of this design intended for liquid samples. Data obtained with this design revealed that a soft extra spacer is necessary to allow for thermal contraction of the sample in the axial direction. Model calculations show that flow in the radial direction is hindered by the confined geometry of the cell when the liquid becomes viscous upon cooling. The method is especially well-suited for—but not limited to—glassy materials.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call