Abstract

Piezocomposite design for dedicated ultrasonic imaging applications requires precise homogenization models for predicting the electromechanical characteristics of the new material. Thus, several homogenization models have been developed. As part of this work, we applied several analytical homogenization models for piezocomposite of 2–2 and 1–3 connectivities. To validate these analytical models, a comparative study was made between various models and experimental measurements. As a result, these homogenized electromechanical properties are effectively used for the calculation and comparison of electroacoustic response for typical transducers aimed at ultrasound imaging applications. An optimal design of transducer aimed at ultrasound imaging applications is proposed as a dedicated imaging performance index, elaborated through a trade-off between sensitivity and bandwidth.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.