An interferogram demodulation method based on azimuthal complex spectrum analysis is proposed for achieving picometer-scale accuracy with an optical vortex interferometer (OVI). The OVI uses conjugated p-radial-order Laguerre-Gaussian beams to produce a high-order petal-like interferogram. A camera with a multi-ring pattern written on its sensor is used to convert the interferogram into multiple azimuthal intensity profiles. A phase shift subjected to either uniform surface displacement or axisymmetric non-uniform surface deformation can be retrieved from the complex spectra of the azimuthal intensity profiles at the main frequency components. The experiment verified that the measurement error is 84 pm for a displacement of 10 nm and 0.359 nm for a deformation magnitude of 100 nm. The effect of surface misalignment on the measurement result is also discussed. The proposed method provides an effective and highly accurate method of interferogram demodulation for the OVI and extends the applicability of OVI from uniform surface displacement measurement to axisymmetric non-uniform surface deformation measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.