Conduction system pacing (CSP) has emerged as a means to preserve or restore physiological ventricular activation via pacing at the His bundle or at more distal targets in the conduction system, including the left bundle branch area. This review examines strengths, weaknesses, and clinical applications of CSP performed via these approaches. His bundle pacing (HBP) has been successfully utilized for standard bradyarrhythmia indications and for QRS correction among patients receiving devices for cardiac resynchronization therapy (CRT). Limitations of HBP pacing have included implant complexity and rising pacing thresholds over time. Left bundle branch area pacing (LBBAP) appears to deliver similar physiological benefits with shorter implant times and more stable thresholds. More recently, hybrid systems utilizing HBP or LBBAP in combination with left ventricular leads have been used to treat heart failure (HF) patients, and may be useful in multilevel or mixed conduction blocks. There is growing interest in CSP for bradycardia and HF indications, although high quality data with randomized controlled trials are needed to help guide future treatment paradigms.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.