Abstract
The primary purpose of mechanical ventilation is to decrease work of breathing. Achieving this goal requires that cycling of the ventilator be carefully aligned with the intrinsic rhythm of a patient's respiratory center output. Problems arise at the point of ventilator triggering, post-trigger inflation, and inspiration-expiration switchover. Careful, iterative adjustments of ventilator settings are required to minimize work of breathing. Use of protocols for the selection of ventilator settings can lead to complications (including alveolar overdistention) and risk of death. Because complications are axiomatic to mechanical ventilation, it should be discontinued at the earliest possible time. To shorten ventilator time, the critical step is to screen for weanability through use of weaning predictor tests. Use of T-tube trials circumvents the impossibility of estimating patient work of breathing during pressure support. Before extubation, patients should demonstrate the ability to breathe successfully in the absence of pressure support and positive end-expiratory pressure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.