A novel design of bioreactor G-BNC, in combination with two previously reported designs of bioreactor were used to fabricate three small caliber bacterial nanocellulose (BNC) conduits (G-BNC, S-BNC and D-BNC). They were compared systematically with a clinically-used ePTFE graft. S-BNC possessed a laminated structure, the lowest BNC content, roughest luminal surface and weakest mechanical properties, and so might not be sufficiently strong for use as an artificial blood vessel alone. The D-BNC conduit possessed an unstratified structure with a fiber network that was more dense and the greatest BNC content, providing the strongest mechanical properties. G-BNC possessed a looser network with the smoothest luminal surface and greater hemocompatibility. Following comprehensive evaluation of mechanical properties and performance, we judge that D-BNC and G-BNC should possess greater potential in application as small caliber vascular grafts, however the patency of the three BNC conduits need be further verified in animal studies in vivo.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.