Abstract

To determine whether the energy expenditure recorded with the physical activity monitor SenseWear™ Pro(2) Armband differs from that recorded with indirect calorimetry. Cross-sectional comparison of measures of energy expenditure. A convenient sample recruited from a randomized controlled trial. Twenty-nine pregnant women (24-43 years old). Energy expenditure was recorded with SenseWear™ Pro(2) Armband and a portable oxygen analyzer for 90 minutes while carrying out a variety of activities of different intensities. Energy expenditure recorded with the physical activity monitor SenseWear™ Pro(2) Armband. Comparing energy expenditure during free-living activities, the mean differences and limits of agreements from Bland-Altman plots was -136±343 kJ, giving an underestimation of energy expenditure by 9%. The intraclass correlation coefficient was 0.85 (95% confidence interval 0.71-0.93; p<0.001), giving 85% of the variance explained by differences among individuals. SenseWear™ Pro(2) Armband is a valid measure of energy expenditure during pregnancy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.