Photothermal deflection spectroscopy was applied to the selective detection of iron(II) chelate with ferrozine by its sorption preconcentration on Silufol plates. The linearity range was 1 x 10(-11) - 6 x 10(-8) mol cm(-2) of chelate at the plate surface, which corresponded to 1 x 10(-9) -4 x 10(-6) M of chelate in solution. The limits of detection and quantification are 8 x 10(-12) and 2.5 x 10(-11) mol cm(-2) at the plate from 15 microL of test solution (0.5 nM and 1.5 nM in solution, respectively), and the absolute detection limit is 8 fmol in the whole spot applied to a plate. Characteristics and features of photothermal deflection detection are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.