Abstract

A new method for heart rate monitoring using photoplethysmography (PPG) during physical activities is proposed. It jointly estimates the spectra of PPG signals and simultaneous acceleration signals, utilizing the multiple measurement vector model in sparse signal recovery. Due to a common sparsity constraint on spectral coefficients, the method can easily identify and remove the spectral peaks of motion artifact (MA) in the PPG spectra. Thus, it does not need any extra signal processing modular to remove MA as in some other algorithms. Furthermore, seeking spectral peaks associated with heart rate is simplified. Experimental results on 12 PPG datasets sampled at 25 Hz and recorded during subjects' fast running showed that it had high performance. The average absolute estimation error was 1.28 beat/min and the standard deviation was 2.61 beat/min. These results show that the method has great potential to be used for PPG-based heart rate monitoring in wearable devices for fitness tracking and health monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.