Abstract

To assess the accuracy of ankle brachial pressure index (ABPI) assessed by photoplethysmography (PPG) compared with continuous wave Doppler (CW-Doppler). Ankle brachial pressure index was measured in a standard manner using both PPG and Doppler probes. For PPG-ABPI, a PPG probe was placed on the index finger and great toe, and a microcomputer determined the ABPI. These values were compared with the ABPI measured manually using an 8-MHz Doppler probe. Correlation and agreement between PPG and Doppler ABPI were assessed by Lin's correlation coefficient and Bland-Altman plots. In all, 133 claudicants were assessed. There was a strong correlation between the 2 ABPI methods (beta = .79 and 95% limits of agreement of -0.23 to 0.24). Measuring ABPI automatically using the PPG technique is an effective alternative for Doppler ABPI. PPG-ABPI is completely objective, fast, and accurate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.