Abstract

We aimed to correlate photoplethysmographic parameters with stroke volume in infants with PDA. Photoplethysmography constitutes the optical signal in pulse oximetry. Stroke volume was determined echocardiographically. Pulse transit time, right hand to foot arrival time difference, and relative amplitude were measured from pulse oximeter and ECG waveforms. Photoplethysmographic parameters before and after PDA closure were compared with stroke volume. After PDA closure, pulse transit time to the hand and to the foot were prolonged (54.7 ± 6.7 vs 65.5 ± 9.8 ms, p < 0.001, 82.5 ± 12.8 vs 88.6 ± 10.6 ms, p = 0.03), arrival time difference decreased (27.7 ± 7.6 vs 23.1 ± 5.6 ms, p = 0.021), and relative amplitude decreased (from 2.1 ± 0.7% to 1.5 ± 0.5%, p = 0.003). The time-based photoplethysmographic parameters correlated with stroke volume. Photoplethysmographic waveform parameters are significantly different before and after PDA closure and the time-based parameters correlate well with stroke volume. Monitoring pulse transit time may assist in evaluation for spontaneous PDA closure or response to therapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.